SEARCH

The Products Section >The Nomad Data Logger

Model mV-HR

Nomad data loggerThe mV-HR is a small, two channel, high resolution (16-bit) voltage data logger, provided with an internal temperature sensor. The unit can accommodate inputs of

  • ±50 mV
  • ±100 mV
  • ±200 mV
  • ±400 mV

Models designed for other voltage ranges, 4-20 mA input, and self-contained loggers for a host of environmental parameters are also available. Please contact us for further details.

Supplied with easy to use yet feature-rich software, the loggers can be programmed to record in engineering units, and yield a number of useful reports. Logging files can be exported to spreadsheet programs, for further analysis.

SPECIFICATIONS

External connector
2 pin Switchcraft Plug EN3C2M
Weatherproof
IP66 rating
   
Pinout
1 Positive
2 Negative
Nomad pinout View from front of socket or rear of plug
   
Maximum readable input voltage 500 mV
   
Maximum voltage input without damaging logger 10 V
   
Input leakage current 1 nA
Resolution
Input range Resolution
   ±50mV 0.01 mV
   ±100mV   0.01 mV
   ±200mV 0.01 mV
   ±400mV 0.1 mV
Accuracy ±0.1% of full scale
If the logger is connected to a computer or instrument with either having an earth ground, erroneous readings will result, because of currents produced from the ground connection.
Temperature sensor Thermistor
Linearity ±0.5°F (Over the range 32°F to 158°F)
[±0.3°C (Over the range 0°C to 70°C)]
Repeatability ±0.2°F   [±0.1°C]
Long term stability ±0.2°F   [±0.1°C]
Working and storage temperature -22°C to +158°F   [-30°C to +70°C]
   
Sampling rate Adjustable in one second increments from 1 sample/second to 1 sample/10 hours
Storage capacity
Logging one channel 
   522,240 samples
145 hours with 1 sec logging interval
362 days with 1 min logging interval
4.9 years with 5 min logging interval
Logging two channels
   261,120 samples
72.5 hours with 1 sec logging interval
181 days with 1 min logging interval
2.4 years with 5 min logging interval
   
Alarms
Two independent alarms
Triggered on any combination of six user configurable alarm conditions
Both alarms can be configured to send SMS messages
Alarms can be reset on user configurable reset conditions
Alarms can be visually checked using the software supplied
   
Start modes
Start immediately
Start on date/time
Start on condition (e.g. temperature > 20°C)
Start on trigger (push button on logger)
Stop modes
Stop when memory is full
Stop on date/time
Loop around (continues logging)
Logging modes
Each channel can be set to log any combination of: 
  • Point readings
  • Average reading
  • Maximum reading
  • Minimum reading
   
Battery
User replaceable
Custom 7.2V Lithium battery pack
One to five years depending on usage 
The data is retained in the case of battery failure
Battery status monitored in software
   
Download time 9 minutes 30 seconds at full capacity
   
Case material Type 304 stainless steel tube
   
Screw on end cap Plated brass
   
Weight 5.3 oz   [149 g]
   
Size 0.79 in diameter x 7.28 L   [20mm dia. x 185mm L]

  

 P.O. Box 2496 • Chatsworth, CA 91313-2496
 TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
 FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com

 Home Page
 Products
     • Portable Survey
          Monitoring Applications

     • Continuous Monitoring-
          Single Point

     • Continuous Monitoring-
          Two Point

     • Automation Based Gas
          Detection Systems

     • Continuous Monitoring-
          Three Or More Points

     • Arc-Max® Computerized
          Data Acquisition System

     • The Halimeter® Measure
          Bad Breath Scientifically

     • Inexpensive, Accurate,
          Nomad Data Logger

     • Vikane® Monitor
 Tech Center
     • Common Specifications
     • Ethylene Oxide     
     • FAQ - Instrument
          Troubleshooting

     • Gas Detection
          Applications Primer

     • Gas Detection
          Knowledge Base

     • General Considerations
     • Getting Service
     • How Do I Get Started?
     • Halimeter®
     • Interfering Gas Data
     • Manual Downloads
     • Material Safety Data Sheets
     • PowerPoint Presentations
     • Schematics
     • Sensor Principle of Operation
     • Sensor Response Data
     • Software Downloads
     • Toxnet - HSDB
     • Useful formulae--
          Conversions, etc.

Industry Links
Interscan In The News
     • Magazine Articles
Representatives and
     Distributors Worldwide

Special Interest Items
     • Mike's Comment
          of the Week

     • Cool Site of the Week
Interscan Photo Gallery
 Streaming Media Library
E-mail Us



Sign up here for our newsletter. (Your e-mail address will NEVER be given or sold to any other party)

e-mail
    Subscribe
    Unsubscribe


Copyright ©1996 - 2008
Interscan Corporation.
All rights reserved.

All other trademarks are
the property of their
respective owners.