The mV-HR is a small, two channel, high resolution (16-bit) voltage data logger, provided with an internal temperature sensor. The unit can accommodate inputs of
±50 mV
±100 mV
±200 mV
±400 mV
Models designed for other voltage ranges, 4-20 mA input, and self-contained loggers for a host of environmental parameters are also available. Please contact us for further details.
Supplied with easy to use yet feature-rich software, the loggers can be programmed to record in engineering units, and yield a number of useful reports. Logging files can be exported to spreadsheet programs, for further analysis.
SPECIFICATIONS
External connector
2 pin Switchcraft Plug EN3C2M
Weatherproof
IP66 rating
Pinout
1 Positive
2 Negative
View from front of socket or rear of plug
Maximum readable input voltage
500 mV
Maximum voltage input without damaging logger
10 V
Input leakage current
1 nA
Resolution
Input range
Resolution
±50mV
0.01 mV
±100mV
0.01 mV
±200mV
0.01 mV
±400mV
0.1 mV
Accuracy
±0.1% of full scale
If the logger is connected to a computer or instrument with either having an earth ground, erroneous readings will result, because of currents produced from the ground connection.
Temperature sensor
Thermistor
Linearity
±0.5°F (Over the range 32°F to 158°F)
[±0.3°C (Over the range 0°C to 70°C)]
Repeatability
±0.2°F [±0.1°C]
Long term stability
±0.2°F [±0.1°C]
Working and storage temperature
-22°C to +158°F [-30°C to +70°C]
Sampling rate
Adjustable in one second increments from 1 sample/second to 1 sample/10 hours
Storage capacity
Logging one channel
522,240 samples
145 hours with 1 sec logging interval
362 days with 1 min logging
interval
4.9 years with 5 min logging
interval
Logging two channels
261,120 samples
72.5 hours with 1 sec logging interval
181 days with 1 min logging interval
2.4 years with 5 min logging interval
Alarms
Two independent alarms
Triggered on any combination of six user configurable
alarm conditions
Both alarms can be configured to send SMS messages
Alarms can be reset on user configurable reset conditions
Alarms can be visually checked using the software supplied
Start modes
Start immediately
Start on date/time
Start on condition (e.g. temperature > 20°C)
Start on trigger (push button on logger)
Stop modes
Stop when memory is full
Stop on date/time
Loop around (continues logging)
Logging
modes
Each channel can be set to log any combination of:
Point readings
Average reading
Maximum reading
Minimum reading
Battery
User replaceable
Custom 7.2V Lithium battery pack
One to five years depending on usage
The data is retained in the case of battery failure