Using Common Sense and Science in Expressing Gas Concentrations

Print
Save to list

Go directly to conversion spreadsheets

Early on, in our first chemistry classes, we learn about solutions, and different ways of expressing their concentrations. Since a solution is any homogeneous mixture of two or more substances, gas mixtures are indeed solutions. In the world of gas detection, we tend to be concerned with combustible or toxic gases dissolved in air, where the gas of interest is the solute and air is the solvent.

Although most combustible gas meters read out in percent LEL (lower explosive limit), that is simply a conversion from the basic measured value in percent by volume. Most common toxic gas detection instruments read out in parts-per-million or parts-per-billion—by volume.

If "a" is the volume of the gas of interest, and "b" is the volume of air, then the volume percent of the gas of interest would be this ratio--

a / (a + b) x 100

[At any given temperature and pressure, the percent by weight, or more rigorously mass percent, would be found by multiplying each component in the above equation by its molecular weight, and then calculating the same ratio.]

If we wish to express the concentration in parts-per-million or ppm, the ratio is simply multiplied by 1 million instead of 100.

Since the properties of gases are greatly affected by temperature and pressure, it is only natural that any parameter to express concentration would be a RATIO OF LIKE UNITS—whether

  • Volume of a component / total volume
  • Weight of a component / total weight
  • Number of moles of a component / total moles [The mole fraction, often used in theoretical discussions]

Unfortunately, this concept, which is intuitively obvious to a beginning chemistry student, has been lost on many who should know better.

In the early 1970's, several government agencies, including our EPA and similar groups in other countries, began to express gas concentrations using mixed unit parameters such as micrograms per cubic meter, or milligrams per cubic meter. Of course, these foolishly contrived parameters are a function of temperature and pressure, and require an asterisk denoting the temperature and pressure at which the value was calculated. More often than not, however, this information is lacking.

But, even if it were provided, since real world gas measurements must occur at fluctuating rather than constant temperatures and pressures, an actual user still has to convert back to parts-per-million. Either that, or he can special order a gas detector that internally converts its ppm output to micrograms or milligrams per cubic meter at a specified temperature and pressure. Yes, we have supplied such units.

Interscan does provide conversion data on its website, from these mixed unit parameters to ppm, and from ppm to the mixed unit parameters. These are the most accurate conversion formulas available, and have been widely downloaded. Also provided from the same gateway page are molecular weights for many compounds.

In addition to being completely impractical, the mixed unit concentration parameters have created their own side effects.

Many potential instrument users are unfamiliar with basic chemistry and are thus unaware that these parameters are affected by temperature and pressure. Moreover, when concentration values are reported to us, using these parameters, they are incorrect at least half the time.

In a recent case, a user mistakenly thought that because in AQUEOUS solutions, milligrams per liter is nearly equivalent to parts-per-million—of the dissolved SOLID—the same would be true for gas mixtures.

Using this "logic," he communicated his requirement for monitoring 5 ppm chlorine as 5 milligrams per liter. In actuality, 5 mg / liter of chlorine in air works out to be 1696 ppm (at 20 degrees Celsius and 760 mmHg).

You might ask how this absurd practice got started. I have two theories, and they are not necessarily mutually exclusive:

1.    Once Canada embarked on its conversion to the Metric system in January, 1970, American environmental officials, knowing that their standards would be emulated throughout the world (and the world was becoming all Metric), wanted to put forth a good face. Thus, environmental standards were set out in Metric-LOOKING units such as micrograms per cubic meter and milligrams per cubic meter. In fact, there is nothing "non-Metric" about parts-per-million, as it is a dimensionless parameter!

Therefore, this mission started out as a fool's errand.

 

2.    In occupational health work, regulatory values are promulgated for many non-gas substances, and these are correctly stated as milligrams or micrograms per cubic meter. Perhaps a desire for consistency across all substances, including gases, crept into the proceedings.

Our best advice is to stick with what works: Percent, parts-per-million, and parts-per-billion. And, if you must convert to the mixed unit parameters, use formulas that are accurate!




Related Articles
More on Expressing Gas Concentrations
This is a follow-up to our earlier article on Using Common Sense and Science in Expressing Gas Concentrations , inspired by a real-life adventure of one of our sales...
When is a part-per-billion (ppb) NOT a part-per-billion (ppb)?
As if it weren't bad enough that clueless environmental authorities try to impose their ignorance on the field by using so-called "metric" units of gas concentration,...
Determining the Dew Point of a Gas Sample
"Dew Point" is defined as the temperature to which a given volume of air must be cooled at constant pressure and constant water vapor content in order for saturation to...

 


  

 P.O. Box 2496 • Chatsworth, CA 91313-2496
 TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
 FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com

 Home Page

 Products
     • Arc-Max® Computerized
          Data Acquisition System

     • Automation Based Gas
          Detection Systems

     • Continuous Monitoring-
          Single Point

     • Continuous Monitoring-
          Three Or More Points

     • Continuous Monitoring-
          Two Point

     • Portable Survey
          Monitoring Applications

     • The Halimeter® Measure
          Bad Breath Scientifically

     • TruTrack Data Logger
     • Vikane® Monitor

 Tech Center
     • Common Specifications
     • Ethylene Oxide     
     • FAQ - Instrument
          Troubleshooting

     • Gas Detection
          Applications Primer

     • Gas Detection
          Knowledge Base

     • General Considerations
     • Getting Service
     • How Do I Get Started?
     • Halimeter®
     • Interfering Gas Data
     • Manual Downloads
     • Material Safety Data Sheets
     • PowerPoint Presentations
     • Schematics
     • Sensor Principle of Operation
     • Sensor Response Data
     • Software Downloads
     • Toxnet - HSDB
     • Useful formulae--
          Conversions, etc.


Industry Links

Interscan In The News
     • Magazine Articles

Interscan Photo Gallery

Representatives and
     Distributors Worldwide


Special Interest Items
     • Mike's Comment
          of the Week

     • Cool Site of the Week

 Streaming Media Library

E-mail Us



Sign up here for our newsletter. (Your e-mail address will NEVER be given or sold to any other party)

e-mail
    Subscribe
    Unsubscribe


Copyright ©1996 - 2010
Interscan Corporation.
All rights reserved.

All other trademarks are
the property of their
respective owners.