"Minimum Detectability" is one of those instrumentation terms that is used frequently, but is seldom defined. Indeed, even though you will encounter this term on many data sheets, its definition does not appear in any of the usual learned references, including Process Instrumentation Terminology, ANSI/ISA—51.1—1979(R1993) and Standard Terminology Relating to Sampling and Analysis of Atmospheres, ASTM D 1356 - 05.

However, the ASTM standard does provide us with...

Method Detection Limit: The minimum concentration of an analyte that can be reported with a 99% confidence that the value is above zero, based on a standard deviation of greater than seven replicate measurements of the analyte in the matrix of concern at a concentration near the low standard.

Simplifying this, we can say that "Minimum Detectability" is the lowest concentration of analyte that can be unambiguously discriminated from noise. [Some agencies set a standard that minimum detectability must be at least 2-2.5 times the noise level.] Fair enough, but how can this be utilized in occupational health or process gas detection?

First of all, it is important to note that any data garnered at the level of minimum detectability will not be accurate. For example, in a typical case, the minimum detectability of a particular instrument is given as 1% of full scale, and accuracy is ± 2% of full scale. Thus, for a 0-100 ppm scale, the minimum detectable reading of 1 ppm would actually be 1 ppm ± 2 ppm—hardly a useful measurement.

Similarly, on a digital unit, the minimum detectability of a particular instrument is often given as the least significant digit. On a commonly used 3½ digit meter, for a range of 0-199.9 ppm, this would be 0.1 ppm. In this case, accuracy is specified at ± 2% of reading ± 1 least significant digit. Here, the minimum detectable reading of 0.1 ppm would actually be 0.1 ppm ± 0.002 ppm ± 0.1 ppm. Technically better than the analog example, but still of little value.

Even so, knowing the minimum detectability of an instrument can be helpful in situations when "go/no-go" readings are of interest. Given a properly calibrated instrument, the smallest observable response would be—by definition—the minimum detectable level, and would indicate at least the presence of the analyte in question (any interferences notwithstanding).

Of course, such practices should only be done when instruments with more appropriate sensitivity are not available.




Related Articles
Accuracy
Definitions "Accuracy" as defined in the ANSI/ISA—51.1—1979 (R1993) standard Process Instrumentation Terminology is: [T]he degree of conformity of an...
Can the detector be calibrated at one concentration to cover the entire range? Since the response is linear, I think the answer is yes. Please confirm. Thanks.
The answer to your question is YES. You are correct in noting that the response of our instruments is linear. More than that, electrochemical voltametric sensors (such...
Our commentary on OSHA's Small Business Guide for Ethylene Oxide
On 30 April 2009, OSHA released a comprehensive document entitled "Small Business Guide for Ethylene Oxide." This new publication is available as a spiffy pdf...

 


  

 P.O. Box 2496 • Chatsworth, CA 91313-2496
 TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
 FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com

 Home Page

 Products
     • Arc-Max® Computerized
          Data Acquisition System

     • Automation Based Gas
          Detection Systems

     • Continuous Monitoring-
          Single Point

     • Continuous Monitoring-
          Three Or More Points

     • Continuous Monitoring-
          Two Point

     • Portable Survey
          Monitoring Applications

     • The Halimeter® Measure
          Bad Breath Scientifically

     • TruTrack Data Logger
     • Vikane® Monitor

 Tech Center
     • Common Specifications
     • Ethylene Oxide     
     • FAQ - Instrument
          Troubleshooting

     • Gas Detection
          Applications Primer

     • Gas Detection
          Knowledge Base

     • General Considerations
     • Getting Service
     • How Do I Get Started?
     • Halimeter®
     • Interfering Gas Data
     • Manual Downloads
     • Material Safety Data Sheets
     • PowerPoint Presentations
     • Schematics
     • Sensor Principle of Operation
     • Sensor Response Data
     • Software Downloads
     • Toxnet - HSDB
     • Useful formulae--
          Conversions, etc.


Industry Links

Interscan In The News
     • Magazine Articles

Interscan Photo Gallery

Representatives and
     Distributors Worldwide


Special Interest Items
     • Mike's Comment
          of the Week

     • Cool Site of the Week

 Streaming Media Library

E-mail Us



Sign up here for our newsletter. (Your e-mail address will NEVER be given or sold to any other party)

e-mail
    Subscribe
    Unsubscribe


Copyright ©1996 - 2010
Interscan Corporation.
All rights reserved.

All other trademarks are
the property of their
respective owners.