Can the detector be calibrated at one concentration to cover the entire range? Since the response is linear, I think the answer is yes. Please confirm. Thanks.

Print
Save to list

The answer to your question is YES.

You are correct in noting that the response of our instruments is linear. More than that, electrochemical voltametric sensors (such as we use) are inherently linear, with no electronic compensation required.

However, for optimum accuracy in most applications, it is best to calibrate the instruments at a concentration somewhere around 50% of the scale range or higher, if you can. This is because there will always be various sources of error beyond the sensor electrochemistry, and prudent analytical technique would frown on calibrating an instrument at 1% of the scale range for readings at 90% of the scale.

Furthermore, for optimum accuracy, if for some reason you must operate toward the bottom of the range, ideally you should calibrate close to this level.

Saying that, for most customers, calibration at 50% of the range or higher should be quite satisfactory.

For more information, please consult our articles on accuracy and minimum detectability.




Related Articles
Minimum Detectability
"Minimum Detectability" is one of those instrumentation terms that is used frequently, but is seldom defined. Indeed, even though you will encounter this term on many data...
Accuracy
Definitions "Accuracy" as defined in the ANSI/ISA—51.1—1979 (R1993) standard Process Instrumentation Terminology is: [T]he degree of conformity of an...
In Search of Zero
As discussed in the Calibration Basics Knowledge Base article, Interscan's gas analyzers, and virtually all other direct-reading gas analyzers are not absolute methods....

 


  

 P.O. Box 2496 • Chatsworth, CA 91313-2496
 TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
 FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com

 Home Page

 Products
     • Arc-Max® Computerized
          Data Acquisition System

     • Automation Based Gas
          Detection Systems

     • Continuous Monitoring-
          Single Point

     • Continuous Monitoring-
          Three Or More Points

     • Continuous Monitoring-
          Two Point

     • Portable Survey
          Monitoring Applications

     • The Halimeter® Measure
          Bad Breath Scientifically

     • TruTrack Data Logger
     • Vikane® Monitor

 Tech Center
     • Common Specifications
     • Ethylene Oxide     
     • FAQ - Instrument
          Troubleshooting

     • Gas Detection
          Applications Primer

     • Gas Detection
          Knowledge Base

     • General Considerations
     • Getting Service
     • How Do I Get Started?
     • Halimeter®
     • Interfering Gas Data
     • Manual Downloads
     • Material Safety Data Sheets
     • PowerPoint Presentations
     • Schematics
     • Sensor Principle of Operation
     • Sensor Response Data
     • Software Downloads
     • Toxnet - HSDB
     • Useful formulae--
          Conversions, etc.


Industry Links

Interscan In The News
     • Magazine Articles

Interscan Photo Gallery

Representatives and
     Distributors Worldwide


Special Interest Items
     • Mike's Comment
          of the Week

     • Cool Site of the Week

 Streaming Media Library

E-mail Us



Sign up here for our newsletter. (Your e-mail address will NEVER be given or sold to any other party)

e-mail
    Subscribe
    Unsubscribe


Copyright ©1996 - 2010
Interscan Corporation.
All rights reserved.

All other trademarks are
the property of their
respective owners.